Two identical, laboratory-certified acoustic calibrators can disagree by over a tenth of a decibel before even leaving for the field.

29:59 Transcript

0:00
0:00
Episode Details
Episode ID
MWP-6964
Published
Duration
29:59
Audio
Direct link
Pipeline
V5.36
TTS Engine
omnivoice
Script Writing Agent
GLM-5.3

AI-Generated Content: This podcast is created using AI personas. Please verify any important information independently.

The prompt

The prompt that started this episode, as sent in (production notes for the hosts removed).

I'd like an episode about calibrated microphones and professional sound measurement, in the context of noise pollution and noise complaints. It belongs with the episodes we've done on digital evidence: a decibel reading is only as good as your ability to show that the instrument was accurate on the day, and that's the first thing the other side will attack.

We've talked before about consumer measurement mics and home-built noise monitors. This time I'm interested in the professional end: the acoustic consultant or environmental health officer whose measurements have to stand up when they testify in court. What are they actually using, which are the main calibrated microphones and sound level meters at that level, and which standards govern them? And what is the calibration process: what happens in the lab, what happens in the field, and what makes the resulting measurement defensible?

Legal battles over noise nuisance turn far less on the sound level meter itself than on the small acoustic calibrator pushed onto the microphone before taking a reading. While meters built to IEC 61672-1 must deliver reliable, time-stamped evidence, the field calibrator provides the prime traceability route back to national measurement standards. Yet this custody chain has notable vulnerabilities. Field calibrators drift, and as UKAS guidance LAB 23 explicitly notes, instruments frequently face harsh temperature extremes when stored in consultants' vehicles. Even worse, routine laboratory testing requires as-received checks before recalibration. If a microphone drifts significantly, the resulting calibration certificate reaches backward in time, immediately casting doubt on months of prior legal evidence.

Sources

What the research for this episode read before the script was written. Primary sources first.

  1. UKAS, LAB 23 Traceability for equipment used in acoustical testing, Edition 3, January 2023 primary retrieved 2026-11-24 .
  2. UKAS, TPS 49 Guidance on the Interpretation of IEC 61672 in the Periodic Testing of Sound Level Meters primary retrieved 2026-11-24 .
  3. IEC Webstore, IEC 61672-3:2013 Electroacoustics primary Sound level meters, Part 3: Periodic tests, retrieved 2026-11-24 .
  4. BSI Knowledge, BS EN IEC 60942:2018 Electroacoustics primary Sound calibrators, retrieved 2026-11-24 (standard dated 31 March 2018) .
  5. legislation.gov.uk, Environmental Protection Act 1990, section 80 primary retrieved 2026-11-24 .
  6. GOV.UK, Noise and vibration management: environmental permits primary retrieved 2026-11-24 .
  7. Institute of Acoustics, *Acoustics Bulletin*, May/June 2022 Sound level meters calibration, what's it all about? Part 2, Ian Campbell, Campbell Associates Ltd, retrieved 2026-11-24 (published May/June 2022).
  8. Svantek, Neighbor noise: sound level from residential premises retrieved 2026-11-24.
  9. NTI Audio, Noise Nuisance Recording System (NNR) retrieved 2026-11-24.
  10. Geonoise Thailand, Acoustic expert witness evidence that stands up in court, retrieved 2026-11-24.
  11. eNoise Control, Acoustic Consultant Expert Witness retrieved 2026-11-24.
  12. Association of Noise Consultants, BS 4142 Technical Note v1.0, March 2020 retrieved 2026-11-24.
  13. LexisNexis / Halsbury's Laws of England, Vol. 78 (2025), 205. Defences retrieved 2026-11-24.

Context

International in scope. Likely terms: IEC 61672 (sound level meters, Class 1 and Class 2; part 3 covers periodic tests), IEC 60942:2017 (sound calibrators, classes LS, 1 and 2), IEC 61094 (measurement microphones). Check current editions, calibration intervals and product names rather than relying on memory. Accuracy of the instrument and validity of the whole measurement (method, position, background noise, weather, the expert's competence) are separate matters. Standards and admissibility rules vary by country; say which jurisdiction any rule belongs to.

Mentions

  • Brüel & Kjær Acoustic measurement and noise monitoring equipment
  • BS 4142 Standard for rating industrial noise affecting residents
  • Campbell Associates Acoustic consultancy, Ian Campbell's calibration article
  • Geonoise Acoustic consultancy marketing transparent measurement uncertainty
  • IEC 60942 Sound calibrator standard with three classes
  • IEC 61672-1 Sound level meter specification standard
  • ISO 17025 Laboratory competence standard for testing and calibration
  • NTI Audio Makers of XL3 noise nuisance system
  • Svantek Professional sound level meter manufacturer
  • UKAS LAB 23 UK guidance on acoustic calibration traceability

Downloads

Episode Audio

Download the full episode as an MP3 file

Download MP3
Transcript (TXT)

Plain text transcript file

Episode Book (PDF)

The episode's record — date, duration, models, sources — with the full transcript

#6760: Calibrated Microphones in Court

Corn
The coverage of noise complaints almost always starts with the meter. Which meter, how much, is it Class 1, did it have a certificate. That's the wrong end of the telescope.
Herman
It is, actually, and it's a nice inversion, because the meter is the easy part. The hard part, the part that decides whether a reading survives a cross-examination, is the little cylinder you push onto the microphone before you start measuring. The calibrator. Nobody writes headlines about the calibrator.
Corn
And yet everybody in the field will tell you it's the first thing a barrister asks about.
Herman
Which tells you something about where the fear lives. The meter is a purchase. The calibrator is a habit. And habits are what fail.
Corn
Daniel's prompt goes exactly there. He wants the professional end of noise measurement, the acoustic consultant or the environmental health officer whose number has to stand up in court. What kit they're actually using, which standards govern it, what the lab does to it, what the field check does, and why the first thing the other side attacks is the claim that the instrument was accurate on the day.
Herman
And he frames it as the acoustic cousin of the evidence episodes we've done. We did one on watermarking and deepfakes a while back, and the parallel holds up: a hash proves a file hasn't changed, a calibration certificate proves an instrument was right when the lab saw it. In both cases the proof is a chain, and in both cases the interesting question is what the chain doesn't reach.
Corn
With the hash, the gap is between the file and the world. With the meter, the gap is between the lab bench and the van.
Herman
The van is going to come up a lot.
Corn
So this is chain-of-custody, except the custody is acoustic. Let's start with the kit, since that's where everyone else starts, and get it out of the way.
Herman
The professional baseline is a Class 1 sound level meter built to IEC 61672-1. That's the specification standard, second edition, 2013, dual-numbered in Britain as BS EN 61672-1. Class 1 means accuracy typically within about one decibel across a frequency range of roughly sixteen hertz to sixteen kilohertz. Class 2 is the general-purpose tier, looser tolerances, and it's what most non-professional work runs on.
Corn
One decibel. That sounds tight until you find out how the meter proves it.
Herman
That's exactly the tension. Hold that thought, because the numbers are worth sitting with. The kit list at the professional tier is a market of several rather than one winner: Svantek, NTI Audio with their XL3 in the Noise Nuisance Recording System, Brüel and Kjær, the old Danish name in the field. No single meter every consultant uses. What the experts do converge on is features, not brands. They want audio recording and data logging, because a reading that's time-stamped and tied to a specific event, a specific fifteen minutes of a specific night, is worth far more in court than a twenty-four-hour average.
Corn
Because an average is anonymous. A log with a timestamp is a witness.
Herman
Right, you can put a timestamp next to a log entry from the complainant saying the fan started at eleven forty. That pairing is where nuisance cases are won. The NTI system is even pitched explicitly at legal proceedings, eviction notices, licensing disputes. Traceable or UKAS calibration, compliant storage of the recordings. The whole product is built around the idea that the recording ends up in front of a magistrate.
Corn
Which is a strange thing to shop for, when you think about it. You're not buying a microphone, you're buying a future exhibit.
Herman
That's actually the correct way to think about procurement at this tier. The question isn't "what's the most accurate meter I can afford," it's "what will survive being handed to a stranger who wants it to be wrong."
Corn
And the answer to that is partly the hardware and partly the paperwork.
Herman
Mostly the paperwork, if we're honest. The standards stack behind all this, before we get into the lab. There's the meter standard, the calibrator standard, and then the laboratory standard that ties them to something national.
Corn
Three layers, and each one references the others. Walk me through how they interlock, because I think people imagine these as independent documents.
Herman
They're anything but. IEC 61672 part one defines the meter classes. Part three, 2013, covers periodic testing, and its stated purpose is almost touching in its modesty, to ensure periodic testing is performed in a consistent manner by all laboratories. Then BS EN IEC 60942 is the calibrator standard, and that one has three classes of its own. LS for laboratory standard, which is normally used only in the lab, and classes one and two for field use.
Corn
The calibrator has classes. The thing that checks the meter is itself classified.
Herman
It's standards all the way down, and each layer is slightly softer than the one above it. Then the top of the stack is ISO 17025, the competence standard for testing and calibration laboratories, and in the UK, UKAS turns that into practice through a guidance document called LAB 23, which is where the traceability obligations actually bite. Current edition is the third, January 2023.
Corn
So the meter standard tells you what a good meter is, the calibrator standard tells you what a good checker is, and 17025 tells you who's allowed to say so.
Herman
That's a fair compression. And the reason it's a stack rather than a single document is that each layer is certifying a different kind of claim. The meter standard certifies design. The periodic testing standard certifies condition. The lab standard certifies the certifier.
Corn
That's the kit and the paperwork. Now the lab. What actually happens when a consultant's meter goes in for calibration?
Herman
This is where it gets good, because there's a beautiful worked example from Ian Campbell at Campbell Associates, writing in the Institute of Acoustics Bulletin. The first thing that surprises people is the shipping requirement. The meter must arrive with its own calibrator. Not any calibrator. The make and model specified in the meter's manual, or a suitable equivalent, conforming to the calibrator standard and of the same accuracy class as the meter. Class one calibrator for a Class one meter.
Corn
It's a matched pair. The lab calibrates the marriage, not just the instrument.
Herman
That's a better phrase than the standard uses. And the associated calibrator, the one the consultant carries in the van, has to have been independently calibrated within the previous twelve months. Campbell's line on why this matters is the whole spine of this episode. He calls the sound calibrator the prime traceability route to national standards. That little cylinder is the physical link between the meter in the van and the national acoustic standard. If the cylinder is wrong, everything downstream of it inherits the error.
Corn
Say more about what "prime traceability route" actually means in practice. Because I think people hear "traceable to national standards" and imagine the meter itself being carried to a vault somewhere.
Herman
Nothing so romantic. The national standard lives at the top in a national metrology institute, and it reaches your field measurement through a chain of comparisons, each one a physical device being measured against the one above it. The lab's reference calibrator was measured against something that was measured against the standard. Your field calibrator gets measured by the lab. Your meter gets checked by your field calibrator every morning. Every link in that chain is a physical object sitting in a room with a temperature, and the link closest to the actual measurement, the last one, is the cylinder in the van.
Corn
And how well is the cylinder known?
Herman
Within a tolerance of plus or minus 0.25 to 0.4 decibels at one kilohertz, depending on which edition of the standard you're working to. Compare that with the meter's own one decibel across the whole band and it looks fine, but the point is the calibrator's certificate can be up to a year old, and the calibrator drifts too. UKAS LAB 23 says it plainly, sound level meters used on site and regularly left in a vehicle may suffer temperature extremes, and the output level of sound calibrators can drift over time.
Corn
The van. The weak link in the national measurement chain lives in a van.
Herman
The van is named in the guidance. I find that delightful. A standards document written by metrologists, people who think in fractions of a decibel, pausing to say: by the way, the instrument lives in a vehicle, and vehicles get hot.
Corn
It's the most honest sentence in the whole stack. Now the worked example.
Herman
Campbell walks through a real report. Manufacturer's reference level is 94.0 decibels at one kilohertz. The lab's own reference calibrator, a Brüel and Kjær 4231, produces 93.93. Then the consultant's calibrator, also a 4231, a different serial number, produces 94.10. You apply environmental corrections, mostly barometric pressure, and other corrections covering things like the difference between pressure and free-field response, and you land on a final statement of level of 93.93 decibels.
Corn
Which is the number the consultant uses in the field every morning for the next two years.
Herman
That's the value to use when the associated calibrator is used for field calibration settings, yes. And the difference between the two calibrators, the lab's and the consultant's, was 0.13 decibels, which Campbell calls well within the expanded uncertainty of the two calibrators. All is well at the start of the calibration. But notice what "well" means there. Two nominally identical instruments, same model, both certified, disagreeing by more than a tenth of a decibel, and that's the good outcome.
Corn
And that's before anything has gone wrong. That's the system working.
Herman
Nobody has dropped anything. Nobody's certificate has lapsed. Two instruments that agree to a tenth of a decibel is the celebration scenario.
Corn
The proof chain has a soft link, and the soft link is the thing doing the proving. What does the lab do when it's not well?
Herman
The before-and-after check is the quietly brutal part. The meter is measured on receipt, before anyone adjusts anything, and measured again at the end. If those two differ significantly, it may indicate problems with measurements made with this meter prior to the calibration. The certificate can reach backwards in time. You send your meter in for a routine check and the lab tells you your last six months of evidence is suspect.
Corn
That's a discovery a consultant never wants to make on a Tuesday.
Herman
And it's not hypothetical. Imagine the sequence. You take measurements in March for a nuisance case. The case is proceeding. In June the meter goes in for its routine two-year calibration, and the as-received check shows the microphone was already down half a decibel in the band that matters. Every reading since the last calibration is now in question, including the one the entire case rests on.
Corn
And the consultant has to disclose that. Presumably the other side gets the certificate.
Herman
The certificate is a discoverable document, yes. Whether they read it closely enough to spot the implication is another matter, but you can't build a defence on hoping your opponent skims.
Corn
There's a drift story in the same article, isn't there.
Herman
There is, and I want to spend a minute on it, because it's the best illustration I've seen of where this whole system almost fails. Campbell shows a six-year calibration history of a half-inch measurement microphone, full calibration every two years. The first four years, fine, drift almost within the lab's own uncertainty. Then the last two years, the mid-frequency response starts to move. The four-kilohertz point comes within 0.24 decibels of the acceptance limit.
Corn
A quarter of a decibel from being out of spec, on a professional microphone.
Herman
And here's the trap. The one-kilohertz check frequency, the frequency every field check uses, fell by only 0.2 decibels over the whole six years. Passing, comfortably. Meanwhile the microphone had gone measurably non-linear between two and ten kilohertz, and it would under-report levels in that band. If your noise complaint is a tonal fan or a whine sitting at four kilohertz, the field check says the meter is fine and the reading is wrong.
Corn
So the single point that's cheap to check is the one point that stays healthy.
Herman
That's the cruel design of it. Microphones don't fail uniformly. And there's a second layer, because even the lab's own data can hide it. Campbell notes that if you only look at the mandated octave resolution, your eye goes to the eight-kilohertz point, where there seems to be a full decibel of headroom to the limit. But with measured microphone data rather than nominal data, the problem is visibly at four kilohertz, with 0.2 decibels to spare.
Corn
The nominal data shows headroom where the real microphone has none.
Herman
Nominal is the manufacturer saying "this model should behave like this." Measured is this serial number, on this day. The gap between those two claims is exactly where the four-kilohertz problem lives.
Corn
It's like the difference between a car's published fuel economy and what your actual car does.
Herman
That's not a bad analogy at all, except the consequence here is that your fuel economy reading becomes evidence in court.
Corn
There's also a negotiation hiding inside "passing," from what you've said.
Herman
The standards permit calibration offsets to centre a marginal instrument in its acceptance band. In the example, another half decibel could be added to the recommended calibration level. Campbell is careful about it, he says this is more common with Class 2 meters, but the key point is that the user must know if such offsets have been used. A meter that passed may have passed with a deliberate nudge.
Corn
A pass is sometimes a settlement, not a verdict.
Herman
And what happened to that drifting microphone is a nice coda. The lab couldn't say whether the change was drift or damage. They laid out the facts to the client, and the client decided to retire the microphone rather than accept a recommended six-month recalibration interval for the kit. Somebody looked at a perfectly functional, expensive, certified microphone and chose to stop trusting it.
Corn
Which is the rational move, and also the one no standard can compel. The standard can say "this is within limits." It can't say "but you should sleep at night." Herman, can I test something on you? You've calibrated the meter, the pair is certified, the certificate is fresh. How much of the reading's defensibility is decided before the consultant leaves the lab?
Herman
Almost none of it, really. The certificate covers the configuration the lab saw. The moment the consultant does something physically ordinary, bolts on weather protection for a long-term deployment, the acoustic path changes. Campbell is explicit that the acoustic effect of front-end accessories can impinge on the results, and if the accessory is supplied at calibration it's assumed it will be used. Swap the microphone capsule and you're in a gray zone. Swap the preamplifier for a different length and the case reflection data no longer applies.
Corn
So the legal question is never "was it calibrated."
Herman
It's whether it was calibrated for this configuration. UKAS LAB 23 has a checklist for microphone replacement, and it reads like a lawyer wrote it. The microphone must be one specified in the meter's manual and maintain the required class. It must have a current and valid calibration certificate. Its sensitivity must be within the manufacturer's normal tolerance limits. And the effect on measurement uncertainty must be evaluated with the meter still conforming to the standard. Miss any one of those and the verification is invalid, and LAB 23 says so in almost exactly those words.
Corn
A microphone that can no longer provide consistent results invalidates the whole verification, not just its own reading.
Herman
The whole chain, retroactively. And that is exactly why the calibrator is the part everyone forgets to doubt.
Hilbert
Hilbert's Mid-Show Haiku. To the Calibrator. Small grey cylinder. You are the only one here who has been checked this year. Checked.
Corn
I'd hang that in the lab.
Herman
Under the certificate, probably. The field side is where consultants actually live, though. LAB 23 names the good practice, checks and setting up before use, a calibrator on the sound level meter before use. Most professionals bracket the session, calibrate before, calibrate after, so the reading is fenced by two known points. The intervals themselves are set by risk rather than one number, uncertainty required, past history, frequency of use, environmental conditions, manufacturer recommendations.
Corn
Reported maximums?
Herman
The figures that get cited are one year for the calibrator, two for the meter, though I'd treat those as reported rather than gospel, since the interval guidance is explicitly a risk exercise. If your meter rides in a hot van through a Jerusalem August, your risk calculation is not the document's risk calculation.
Corn
No. Ours is a shelf by the window, which is its own microclimate. The field check brackets the session at one kilohertz. You've told me what it can't see. What is it actually doing for the consultant?
Herman
It's proving continuity. It says the system that was verified in the lab is the same system, responding the same way, today. It catches gross failures, a cable gone bad, a capsule that took a knock, a moisture problem. What it cannot do is catch the two-to-ten-kilohertz non-linearity, because it only ever asks the one question, at the one frequency, where the answer is reassuring.
Corn
It's a smoke detector, not a medical. It tells you the house isn't on fire. It doesn't tell you the pipes are furring up.
Herman
That's precisely the division of labour, and it's why the two-year lab visit exists at all. The field check is cheap, fast, and shallow. The lab visit is expensive, slow, and deep. The system assumes you do both and knows that neither one alone covers the failure modes that matter.
Corn
And then, above the instrument entirely, there's the method. Where you put the microphone, the wind, the rain, the background. UK guidance says don't monitor above five metres per second of wind or in rain.
Herman
Position and weather. Herman's two favourite subjects after gain staging.
Corn
You said that, not me.
Herman
Position matters more than equipment, I've said it before and a room that flatters a signal path will lie to you as soon as look at you. But the legal layer is the part I find strange, and I want your read on it, because you're the one who argues that adversarial pressure is what produces standards. The courtroom doesn't ask what the reading was. Under section eighty of the Environmental Protection Act, in proceedings for failing to comply with an abatement notice, there's a defence to prove that the best practicable means were used to prevent or counteract the nuisance. Best practicable means, reasonably practicable having regard to local conditions and circumstances, technical knowledge, financial implications.
Corn
That's not a measurement. That's a judgment about somebody's budget.
Herman
It's a question about what a reasonable operator could have spent. And BS 4142, the standard that gets cited in half these disputes, explicitly refuses to decide nuisance. The Association of Noise Consultants is blunt about it, nuisance is determined case by case by the courts and the relevant considerations are far broader than those covered by the standard. So you have this calibrated, uncertainty-stated measurement feeding a legal test that isn't a measurement at all.
Corn
The instrument chain gets you to a number. Then the number enters a world where the question is whether spending more money was reasonable.
Herman
That gap is where expert witnesses earn their fees. The uncertainty handling is the hinge. The ANC argues that the consideration of uncertainty is a fluid process which ought to factor into every stage of the assessment and should be revised with any new information. Their own illustration is that plus or minus two decibels may matter little where the rating level is fifteen decibels below background, but greatly where it's two decibels above. Same uncertainty, opposite consequences.
Corn
Two decibels of slop that means everything or nothing depending on where you're standing relative to the background.
Herman
Context turns the number into evidence. The regulator's own instruction backs that, operators must identify and minimise the sources of uncertainty associated with any measurement or prediction. Not eliminate, minimise. The system is honest about its own limits, which I respect.
Corn
Here's the thing I kept waiting for, though. Daniel's framing is that proving the instrument was accurate is the first thing the other side will attack. Everyone in the industry says that. Geonoise markets exactly it, we state our methodology and measurement uncertainty openly, so the evidence can withstand scrutiny from the other side, not just support your position. Where's the case law?
Herman
I looked for a reported case where a decibel reading was excluded or discredited specifically because of a lapsed or missing calibration certificate, and I can't find one. The attack vector is real as professional consensus and as a commercial selling point, but I can't point you to a judgment where it landed.
Corn
Everyone rehearses for the attack that never shows up in the record.
Herman
Or the attacks settle. Or they succeed quietly, in cross-examination, and the judgment mentions credibility without naming the certificate. My guess is the attack mostly works by erosion, the expert concedes the uncertainty, the reading carries less weight, and nobody writes a precedent about weight.
Corn
You'd expect the threat alone to discipline practice, same as anything else that gets cross-examined. The consultant who shows up with an in-date chain never has to find out what the attack costs.
Herman
And that's arguably the system working as designed, just not as documented. The standard doesn't have to be enforced in a courtroom to shape behaviour if everyone believes it would be. The belief does the enforcement.
Corn
A deterrence system with no recorded convictions.
Herman
The quietest kind of regulation there is.
Hilbert
Two calibrators, both 4231s, and they disagreed by 0.13. That's the normal case.
Herman
That's the Campbell example, yes.
Hilbert
I rang a man about that this afternoon. While Herman was doing the van speech. Our neighbour has an air source heat pump on the party wall, and for two years I've been telling him it hums. He says it doesn't. So I borrowed the council's kit, you can, if you write to them, a woman called Rina brings it round in a case. I calibrated the meter on the doorstep, before and after, ninety-four decibels at the hum frequency, all square, and I got a reading I liked, so I wrote it on the back of a bill.
Corn
What did the bill say the hum was?
Hilbert
Forty-one. At night, with the fridge off, which I unplugged for the duration and plugged back in after, timing it on the stove clock. The next morning I rang the man again to ask whether the reading would hold up if he disputed it, and he asked me what the calibrator's certificate said. I hadn't asked Rina. I've since asked Rina. She's checking.
Corn
So the one number you liked is currently resting on a certificate nobody's seen.
Hilbert
It's resting on Rina, and Rina has never let me down on a Tuesday. The thing I wanted to know, Herman, is whether the reading would still stand if I'd swapped the windshield on, because I put the foam on halfway through. It was drizzling.
Herman
The windshield changes the acoustic path. If the calibration was done bare and the measurement was taken with foam, that's exactly the configuration gap LAB 23 worries about. It's a small effect at most frequencies, but small is what this whole episode is about.
Hilbert
Then I'll take the foam number off the bill. There were two numbers. I'll keep the bare one.
Corn
The bill is your calibration certificate.
Hilbert
The certificate is in the post, apparently. Forty-one at night is still forty-one. Herman, you spent forty minutes on the soft link before I got a word in, and I now think my heat pump dispute is smaller than I believed it was, which I'll admit once.
Herman
Was the reading before or after midnight?
Hilbert
After. I unplugged the fridge at five past, plugged it back in at twenty past. The stove clock and I have an arrangement.
Corn
Two numbers, one bill, and a witness called Rina. More chain of custody than most consultants manage.
Herman
The foam detail is the part I keep coming back to, honestly. That's the whole configuration problem in one drizzly evening on a doorstep. He did the bracket check right, before and after, and then quietly changed the instrument halfway through because of weather, which is the most reasonable thing in the world to do.
Corn
And the weather check is in the guidance too. Don't measure in rain. He measured in drizzle with a foam hat on. Every clause exists because somebody did exactly that.
Herman
The judgment call he can't escape is what forty-one decibels at night means. Is the hum a nuisance. That's the court's question, not the meter's, and the meter can't even see the tonal character of the thing unless the weighting and the analysis are set up for it. He's got a number. The number doesn't decide anything.
Corn
Which means his neighbour can be entirely sincere and entirely annoyed, and both of them are right about the forty-one.
Herman
The open question this leaves me with is the one Daniel's framing smuggles in. If the calibrator is the weak link, and the field check can't see the failure that matters, how much should we trust any single decibel reading, even a professional one? The chain is only as strong as its softest link, and the soft link is the thing doing the proving, not the thing being proved. The confidence is real, but it's real the way a warranty is real, bounded and dated and attached to a specific configuration.
Corn
Mine too, and Hilbert Flumingtop produces this show, which this week apparently includes doorstep metrology. For more along these lines, there's episode nine, Benchmarking Custom ASR Tools - Beyond The WER; episode eight, Building Your Own Whisper; and episode five, Fine-Tuning ASR For Maximal Usability. This has been My Weird Prompts. If today's episode earned its keep, leave us a review wherever you listen. We'll be back soon.
Herman
See you then.

This episode was generated with AI assistance. Hosts Herman and Corn are AI personalities. It may contain errors: disclaimer · report an error